Product Introduction
Intelligent instrument calibrator for PT CT error and impedance admittance test.It meets the requirements of JJG313-94 current transformer verification regulation and JJG314-94 voltage transformer verification regulation.
Product Characteristics
1、Intelligent transformer calibrator adopts 320 240 lattice liquid crystal, with large visual range and long life backlight, convenient to use
2、Chinese operation interface, beautiful interface
3、The percentage difference Angle difference is displayed in large font, which is convenient for users to observe
4、The intelligent transformer calibrator produced by our company fully meets the requirements of jjg314-1993 and jjg314-1994, and it can be sampled automatically
5、Automatically judge polarity error, change ratio error
6、The upper computer software has the open function of virtual instrument
7、Completely solve the s-level measurement problem
8、Automatic range switching
9、The perfect combination of advanced and unique circuit and DSP technology can completely remove the problem of the circuit instability
10、power dissipation:<15VA(No micro printer);<25VA(microprinter)
11、Harmonic suppression ratio:>40db
12、boundary dimension(Band dimensions):260×350×150mm3
13、weight:6kg
פרמטרים טכניים
| סביבת תפעול של ציוד | temperature:5℃~40℃ | relative humidity:<80%(25℃) |
| altitude:<2500m | ||
| mains frequency:50Hz±0.5Hz | mains input:220V±5V | |
| טווח מדידה | cophase component(%):0.0001~200.0 | distinguishability:0.0001 |
| orthogonal component(part):0.001~700.0 | distinguishability:0.001 | |
| impedance(Ω):0.0001~20.0 | distinguishability:0.0001 | |
| admittance(ms):0.0001~20.0 | distinguishability:0.0001 | |
| טעות מהותית | רכיב cophase | ΔX=±(X×2%+Y×2%)±Dx(Level 1 is optional) |
| מרכיב אורתוגונלי | ΔY=±(X×2%+Y×2%)±Dy(Level 1 is optional) | |
| "X"、"Y"——Display value of instrument | ||
| "Dx,Dy"——Display value of instrument | ||
| Dx=2,Dy=5 | ||
| מחוון חיוג | PETS2 (רמה 1 היא אופציונלית) | |
| טווח עבודה | חַשְׁמַל | (1%~149%)In (In =5A) |
| (5%~149%)In (In =1A) | ||
| מתח | (5%~149%)Un (Un =100V,150V,100V/) | |
| (5%~149%)Un (Un =100V/3) | ||
| עומס עבודה | חַשְׁמַל | TO to TX<0.12Ω cosΦ=1 |
| מתח | a to x<0.25VA (100V) | |
| חיווי שגיאת קוטביות | כאשר זרם ההפעלה (מתח) מגיע ליותר מ-5% מהזרם הנקוב (מתח) והשגיאה עולה על 180%, צריכה להיות חיווי קוטביות. | |
| הערה: אם זה יותר מ-10% מזרם ההפעלה המדורג (מתח), עדיין אין חיווי קוטביות נאות, המעיד על תקלה. נא לא להגביר את הזרם (מתח) כדי למנוע שריפת המכשיר. | ||
| חיווי שגיאה ביחס משתנה | כאשר זרם ההפעלה (מתח) מגיע ליותר מ-5% מהזרם המדורג (מתח) והשגיאה עולה על 30% אך קטנה מ-180%, יש לתת את חיווי השגיאה של היחס המשתנה. | |
| בדיקת בידוד ומתח ו הוראות | terminalTX( | |
| The power socket can withstand the housing1.5kV,1minwithstand voltage | ||